AOX-400 Automatic Organic Halogen Analyzer

AOX-400 Automatic Organic Halogen Analyzer

Nittoseiko's AOX-400 Automatic Organic Halogen Analyzer measures organic halogen in river water, industrial wastewater, and soil via shaking method, column method, and EOX.

Advanced autosamplers provide flexible combination of different modes easily.

Main Features

  • 3-injection mode convertible (shaking method/ column method/ EOX)
  • Advanced autosampler concept is space saving and reliable
  • Faster measurement time
  • Automatic gas flow
  • No gases required except air (same as AOX-200)
  • Longer lifetime of electrolyte by cell cooling system
  • Increase the default number of samples
  • Easy maintenance of pyrolysis tube by sliding turntable

nittoseiko

Measurement Principal

AOX adsorbed to activated carbon is pyrolyzed in a quartz combustion tube. Combustion gases containing hydrogen halides are forced through a dehydrating tube and subsequently adsorbed within the electrolyte of a coulometric titration cell.

Within the titration cell the halogen ions are quantified by an argentometric coulometry. The amount of chlorine is calculated from the quantity of electricity required for the formation of silver ions.

HCI + Ag+ ➔ H+ + AgCI (Titration)
Ag ➔ Ag+ + e- (Electrolysis)

The gas is moved by application of suction pump.

This principle is used to destroy acetic acid vapour by returning the gases through a separate combustion tube within the furnace after passage of titration celI.

diagram

Official Method

  • ISO 9562, DIN EN 1485
    • Water quality-determination of adsorbable organic halogens {AOX/EOX)
  • EPA 9020
    • Total organic halides {AOX-Column method)
    • ICR [EPA 814-8-96-002 for QC]
  • DIN 38414 part 18
    • Sludge and Sediment-Determination of adsorbed organically bound halogens {AOX-Shaking method)

Repeatability

(AOX) Confirmation sample: 4-chlorophenol solution

Concentration (ng/ml) Sample Volume (ml)  Theoretical Value (µg) Recovery (%) RSD (%) n=3 
10 100 1 80 ~ 120 ≤10
100 20 2 95 ~ 105 ≤3

(EOX) Confirmation sample: 4-chlorophenol/hexane solution

Concentration (µg/ml) Sample Volume (µl)  Theoretical Value (µg) Recovery (%) RSD (%) n=5 
5 100 0.5 80 ~ 110 ≤8
50 40 2 85 ~ 105 ≤3

interior


Specifications

Measuring Object

River water, Industrial wastewater, Soil

Sample forms

Solid (AOX: activated carbon adsorbed liquid sample), Liquid (EOX: extraction solution)

Analysis method

Oxidative pyrolysis/Coulometry

Furnace temperature

150~ 1000 °C (Regular use: 950° C)

Sample introduction

Introduction to the open top pyrolysis tube driven by Newton's law

AOX: Extrusion injection of activated carbon adsorbed sample (Column method) Drop frit with activated carbon adsorbed sample (Shaking method)

EOX: Direct injection by syringe

Measurement range

Total organic halogen: 0.1 to 50 µ1g

Sample volume

AOX sample: 50mg or less, EOX sample: 200 µL or less

Measurement time

Within 10 minutes/measurement (At 2 µg sample measurement)

Number of samples

Frit: 30 samples

Column: 60 samples

EOX: 40 samples (Size: 2ml vials with septum)

Gas

Using AIR-200, Air gas is not required.

AOX: Air (800mL/min)

EOX: Air (l00mL/min) + Oxygen (300mL/min)

Operation Condition

15 to 35 °C, 75%RH or less (no condensation)

Power supply

AC 100/115/230/240V, 50/60Hz, 960VA

Dimensions & Weight

AOX main unit: Approx. 460(W) x 420(D) x 570H (H) mm, Approx. 37kg

Sl-400: Approx. 190(W) x 400(D) x 410H(H) mm, Approx. 7kg

AIR-200: Approx. l00(W) x 400(D) x 220H (H) mm, Approx. 5kg

Device control and data processing

PC, Printer: compatible with Windows®

dimensions


PDF