GT-310 Automatic Titrator
High-precision titration automatic buret and the new titration control method enables accurate titration. Up to four units can be connected and measured simultaneously
Main Features
4-channels titration
Combination of up to 4 potentiometric and Karl Fischer titrators can be connected in one Multi-controller with simultaneous measurements.
Automatic pH calibration
Equipped with automatic stability judgement function. Automation from pH calibration to sample measurement is possible when using MST.
Reanalysis function
Reanalysis to change end point detection is now possible.
Inflection point ⇔ set potential ⇔ front intersection ⇔ rear intersection ⇔ V intersection
LIMS connection
Multi-controller transfers measuring data automatically to the network simultaneously.
Data Integrity
GT-310 with integrated data integrity feature, comes with strong and smart support for data protection and management without a dedicated PC. Audit trail function is provided as an option in the data integrity version of the GT-310 software.
Audit trail | (option) |
---|---|
User Management, Login Password, Permission Management | ✓ |
Validation | ✓ |
Backup | USB memory/USB-HDD/networking |
Maintenance management | ✓ |
Five-step certification of measurement results | (option) |
Networking with LIMS | ✓ |
Main Specifications
Type of titration |
Potentiometric titration (acid-base, redox, chelatometric, precipitation) polarization and conductivity titration (optional) |
---|---|
Detection range |
pH: 0~14 -2000mV to +2000mV |
End point detection |
Test titration General titration Petroleum neutralization value-official method/conventional method, etc |
Number of files | Sample name: 99 cases, Blank value: 99 cases, Factor: 99 events,etc |
Buret accuracy | 20ml ± 0.02ml |
Resolution/Minimum drip rate | 0.001ml |
Number of stored data | 9999 cases |
GLP-compliant function | Validation (electrode inspection, injection volume inspection, measurement inspection), history display of pH calibration, user registration function with password, |
Dimensions and weight |
MC: Approx. 245(W) x 160(D) x 215(H) mm STR: Approx. 110(W) x 165(D) x 415(H) mm 5.5kg BRT: Approx. 127(W) x 378(D) x 260(H) mm 1.2kg |
Optional Multi-sample Titrator GT-310MST (Sample Changer Function)
GT-310, with a Multi-controller and burets, GT-310MST performs multi sample / various sample measurement continuously. Both 12 and 24 beaker storage are available.
Number of stored beakers
|
12 or 24 |
---|---|
Beaker size | 100/200 ml beaker (options: 100/200/300 ml tall beaker, 100/200ml disposable beaker, 100 ml Erlenmeyer flask) |
Electrode washing | Automatic washing in dedicated washing pot, Showering, and sinking washing |
Table rotation method | Turntable, rotation driven by DC motor |
Beaker detection | Optical sensor |
Dimensions and weight | MC: Approx. 414(W) x 489(D) x 423(H) mm 15.4kg |